Gamma-Ray Spectroscopy with Germanium Detectors - Response Function

this index of detector performance is defined as ratio of the count in the highest photopeak channel to the count in a typical channel of the Compton continuum associated with that peak

sample of the continuum is to be taken in the relatively flat portion of the distribution lying just to the left of the rise toward the Compton edge

ratio is conventionally quoted for the 1333 keV gamma ray from 60Co