Operating humidity (IEC 60068-2-56).........................

.....10 to 90% RH, noncondensing

Storage humidity (IEC 60068-2-56).........................

.....5 to 95% RH, noncondensing

Maximum altitude........................ Pollution Degree (IEC 60664) .....

..... 2,000 m ..... 2

Shock and Vibration To meet these specifications, you Operating vibration

must panel mount the system.

Random (IEC 60068-2-64).... Sinusoidal (IEC 60068-2-6) ..

.....5 grms, 10 to 500 Hz .....5 g, 10 to 500 Hz

Operating shock (IEC 60068-2-27).........................

.....30 g, 11 ms half sine, 50 g, 3 ms half sine, 18 shocks at 6 orientations

© National Instruments Corp.

19

NI 9476 Operating Instructions and Specifications