Manufacture of sterile medicines – Advanced workshop for SFDA GMP inspectors

Nanjing, November 2009

26 |

Equilibration Time

The equilibration time is the period that elapses between attainment of the minimum specified sterilizing temperature in the chamber (chamber reference temperature - typically in the drain) and attainment of the minimum specified sterilization temperature in the load, as measured by the slowest-to-heat penetration probe.  This period is an indication of the ability to properly condition the load through air removal and load heating.